|
Technical Tips
Reference Electrodes
Electrochemists like reference electrodes
but potentiostats can have a
different opinion. That lovingly crafted jelly filled bridge leading from a
bubbly Lugin to a clogged up SCE placed next to the rotating-disc motor
will give the
instrument something to chew on. Most students have electrolysed cells dry as a
result of a bubble in a Lugin. I know I did.
If the absolute potential is not so important
but you are looking for
changes or rate of change in the current with respect to voltage
then consider another
piece of the working electrode immersed into the electrolyte. Its potential will
probably remain fairly constant during the test
given a simple non-dynamic situation
and
its impedance is most likely very low. High impedance reference electrodes cause
three problems:-
1. DC offset
Slight DC offset due to the finite input current of
the reference electrode terminal - say a 1 K Ohm SCE into a 20 pA input bias current RE
terminal
giving 20 nV offset - is no problem. The situation changes if a multimeter is
connected between RE and WE
which can typically have an input resistance of 10 M Ohm
making the situation orders of magnitude worse. The motto is disconnect
DVMs when
doing tests.
2. AC effects
When using reference electrodes for AC Impedance there
are a couple of effects to watch out for. Firstly the interrelationship between the
resistance of the RE and the input capacitance of the RE terminal gives an effect of low
pass filtering. This is usually acceptable at the typical corrosion frequencies of 20KHz
or less
but gives perhaps a degree of phase shift at the top frequencies. The second problem is
the fact that reference electrodes are themselves electrochemical cells
possessing their
own capacitance
resistance and time constant. With an SCE above 17 kHz people
generally end up measuring the reference electrode.
3. EMI
A method of virtually eliminating electromagnetic interference has been
invented by ACM. Using the ACM fourth electrode the noisiest of cells can be
measured.
An example of our new method
Stability and Time Constants
A perfect potentiostat is unconditionally stable with every cell.
The way all potentiostats are constructed at present involves feeding an analogue signal
back from the reference electrode to the virtual earth of an operational amplifier.
The capacity for introducing phase shifts occurs on the WE; the AE and the RE
these
rendering current design potentiostats liable to oscillation. At ACM we tackle
this using practical technology.
A potentiostat should only be as fast as it needs
to be. This sounds
simple but how fast does it need to be to perform a cyclic sweep at 1 mV/sec?
You can buy potentiostats that let you or your technician perform this essentially DC test
with a frequency response on the potentiostat of 1 MHz and the same on the ZRA (current
measuring stage). This is a recipe for disaster. The software should look ahead to
the forthcoming test and set the appropriate speed. All of ours do this
slowing
down the control amplifier when needed and damping the ZRA. At each current range
the time constant of the ACM ZRA is the same
allowing very smooth current range
changes.
IR Compensation
This paragraph is in the wrong
location: it should be above the preceding
one
because (in)stability follows current interrupt IR compensation
as Brazilians
follow football. Current interrupt IR compensation looks good on paper
but in
practice it can be the biggest cause of unstable potentiostats there is. The best
way to compensate for IR drop is to measure the solution resistance with an AC signal
and then
adjust the output voltage with a simple algorithm. We know that this is much more stable than
current interrupt
having made both. The reason people use current
interruption is one of cost
as it's much cheaper to put a little interrupter in the circuit
than incorporate a full blown AC analyser. Using our own on-board DSP the solution
resistance can be found in 1 second
then used correctly. We found when we made our
current interrupter
prior to using the DSP
we spent all our time answering the phone
over sampling speeds
cell oscillation
cable capacitance and noise. That's the same
noise that creeps into high electrolyte impedance cells - just the sort of cell that IR
compensation is needed for. As for using current interrupt when measuring AC
Impedance...!
AC Impedance
During AC Impedance tests the potentiostat is put into a faster mode.
The compromise between stability and speed is most apparent when measuring high
impedance cells. A few pF of capacitance is needed across the ZRA;
Potentiostat
and
RE to maintain stability. This then determines the maximum operating frequency for
any current range. Our software understands this maximum and sets the appropriate
range. This does not mean that a 100 MOhm cell can then be measured at 100
kHz. The
practical limit for say a 100 MOhm cell is about 10 Hz . All manufacturers are
constrained by the same laws of physics so only systems with impedances of around 100
-1000 Ohms can be measured with any accuracy using a top frequency of 100 kHz.
Using our Paint Buffer and measuring a cell of 1 GOhm the top frequency for
accurate measurement is about 1Hz
the same as other honest manufacturers. We do
however have an interesting device to squeeze more accuracy out of an instrument.
Each of our instruments that incorporate AC (Gill 8
Field Machine
Gill AC
and Gill 12)
undergoes a very extensive calibration procedure over a series of 8 decades of cell load
10 amplitudes and the full range of frequencies to produce a calibration matrix we call
the P.A.I.R. table (Phase and Impedance Reduction). This is then applied to
the raw data to reduce cable and instrument induced errors. The instruments can
easily be recalibrated using any length and type of cable. For example 60 feet of cable up
a power station flue can easily effect the results when looking for dew point corrosion
but not if calibrated with a P.A.I.R. table.
Big Currents
Some people need power
and not just an Amp or two
but tens
even hundreds of
Amps. We make big current systems. They are not cheap but they do work.
Last tip
The last tip is just a note to scientists and technicians who come to
testing for the first time. Don't be afraid to have a play with your system at
first. Try simple jam jar tests with chunks of electrodes taped to the side.
Waggle electrodes about when polarising. Try the effect of widely different sweep
rates
and polarising voltages. Shine lights on the cell
turn on motors
scratch
electrodes and bubble the electrolyte. Try altering cell geometry and generally get
a feel for the orders of magnitude of the system. Then tighten up your experimental
design
not the other way round.
When the design is tightened up
watch out for those bubbles in
Lugin probes; broken connections; crevices between sample and holder; dry SCEs; touching
electrodes and inadvertent ground loops. If your computer is to be left unattended
- remove all games and don't activate a screen saver (your test will cost more than a
monitor) and make sure everything is hard to unplug. Let people know that a test is
underway and sleep soundly.
|